Hydro-Probe XT User Guide HD0538 Rev: 1.6.0 17
2.1.2
Bin wall mounting
The sensor can be placed horizontally in the bin wall, or if the space is limited, angled
down to 45° as shown, using the standard mounting sleeve.
•
The sensor should be placed in the centre of the widest side of the bin and, where
possible, mounted on the opposite side to any vibrators (where fitted).
•
Ensure the sensor ceramic is not mounted closer than 150mm to any metal work.
•
Ensure the sensor does not obstruct the door-opening.
•
Ensure the ceramic face-plate is in the main flow of the material.
Figure 7: Mounting the Hydro-Probe XT in the bin wall
If the sensor does not reach the main flow of material, then an extension mounting sleeve
should be used, as shown in Figure 8.
Figure 8: Mounting the Hydro-Probe XT in large bins
Standard Mounting
Sleeve
(part number: 0025)
Extension
Mounting Sleeve
(part number
0026)
Содержание Hydro-Probe XT
Страница 4: ...4 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 6: ...6 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 8: ...8 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 10: ...10 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 14: ...Chapter 1 Introduction 14 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 42: ...Chapter 4 Configuration 42 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 52: ...Chapter 6 Optimising Sensor and Process Performance 52 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 58: ...Chapter 7 Sensor diagnostics 58 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 60: ...Chapter 8 Technical specification 60 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 66: ...Appendix B Document Cross Reference 66 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...