Hydro-View / Hydro-Hub User Guide HD0864 Rev 1.0.0 61
The corresponding moisture % associated with the Unscaled value is added manually in the
Moisture % column. The required moisture and Unscaled values can be added to the
calibration by clicking the include column for each point (Figure 96).
Figure 96: Moisture % Added to Chart
Click Calculate to add the selected points to the calibration graph.
Figure 97: Calculate Calibration
Once the calculation is completed the calibration coefficients are updated and displayed on
the information tab (Figure 98). An R² value for each measurement modes coefficients is
also included. The R² can be used to indicate how close the data in the calibration is to the
calculated best fit line. An ideal calibration where the best fit line passes through every point
will result in an R² of 1.
Figure 98: Updated Calibration Coefficients
Click the Calibration graph tab to display the graph (Figure 99).
Содержание Hydro-Hub
Страница 4: ...4 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 6: ...6 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 8: ...8 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 12: ...12 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 14: ...Chapter 1 Introduction 14 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 20: ...Chapter 2 Mechanical Installation 20 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 38: ...Chapter 4 System Overview 38 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 44: ...Chapter 5 Remote Access 44 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 70: ...Chapter 7 Calibration 70 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 72: ...Appendix A Quick Start Rules 72 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...
Страница 74: ...Appendix B Document Cross Reference 74 Hydro View Hydro Hub User Guide HD0864 Rev 1 0 0 ...