25
Water Tank Temperature Alarm Testing
a. Remove the thermocouple located at the heater outlet (control water temperature).
b. Place in a bath of room temperature water, allow the sensor to reach equilibrium and check the low
water temperature alarm will activate.
c. To check the high water temperature alarm, boil water in a separate container. Place the
thermocouple in the nearly boiling water (approximately 210ºF / 99ºC) to verify that the high water
temperature alarm will activate.
d. Re-install the thermocouple into the piping system.
Superheat Alarm Testing
The superheat alarm can be tested by placing the chlorine gas temperature probe in ice during the
pressure switch alarm testing.
Cathodic Protection Testing
Disconnect the red anode wires. The current value should then drop to zero.
1.3 Component Preventative Maintenance
All rupture discs should be replaced on a yearly basis. O-rings and gaskets should also be replaced
yearly or sooner if they appear cracked/damaged in any way. Before proceeding to perform any test-
ing or maintenance ensure that there is no chlorine in the system and that the chlorine inlet piping is
shut off. Please refer to Section III.4.2 for system shut down.
Water Recirculating Pump
The water recirculating pump has a replaceable internal cartridge, which should be replaced if the
pump is noisy, leaking, and/or not circulating water.
Magnesium Anode Replacement
Based on normal operating conditions, the four (4) sacrificial magnesium anodes will need to be re-
placed yearly or more often if necessary. This is a very important step as it prevents the chlorine pres-
sure chamber and all other system components from corrosion.
a. Remove wires connected to the anodes.
b. Remove anodes.
c. Install new anodes into the respective locations and attach the anode wires.
2. 5-Year Chlorine Pressure Chamber Cleaning and Inspection
This step should be conducted at least every five years or if signs of corrosion exist.
a. Complete Section IV.1.1 chlorine pressure chamber cleaning (perform steps a through h only).
b. Weigh the inner chlorine chamber assembly (less the process instrumentation, water tank bolts,
anodes and fittings, and inlet/outlet pipe.) and record the weight.
c. If the weight of the inner pressure chamber is 10 pounds less than the original weight, or corrosion
pits exceeds more than
1
⁄
8
" deep anywhere on the chamber, it must be replaced. Original weight is
stamped on the pressure chamber and also stored with Hydro Instruments for future reference.
d. Complete Section IV.1 steps i through n.
Содержание VPH-10000 Series
Страница 1: ...1 Series VPH 10000 Vaporizer Instruction and Operation Manual VPH 10000 Rev 7 21 14 ...
Страница 11: ...11 Figure 2 Main Controller Wiring Figure 2 Main Controller Wiring ...
Страница 12: ...12 Figure 3 Relay Output Box Figure 3 Relay Output Box ...
Страница 32: ...32 Figure 8 Chlorine Vapor Pressure Curve Figure 6 Chlorine Vapor Pressure Curve Vapor Pressure ...
Страница 34: ...34 Figure 10 Ammonia Vapor Pressure Curve Figure 8 Ammonia Vapor Pressure Curve Vapor Pressure Min Superheat ...