I. INTRODUCTION
The Hydro Instruments TH-4000 turbidimeter uses a pre-scaled inline sensor that measures
turbidity in water using a white LED as the excitation light source and by measuring the
scattered light at a 90-degree angle with respect to the excitation beam. The
fl
uidic and
optical arrangement of the sensor is designed to overcome many shortcomings associated
with other inline turbidimeters. It can be easily inserted into the
fl
ow cell tee with a
compression
fi
tting designed to ensure correct positioning of the Series sensor in the water
stream. The sensor has a short channel that can be easily cleaned and calibrated.
Speci
fi
cations
Speci
fi
cations
TSH-10
TSH-100
TSH-1K
TSH-10K
Turbidity Range:
0-10 NTU
0-100 NTU
0-1,000 NTU
0-10,000 NTU
Resolution:
0.05 NTU
0.10 NTU
1.00 NTU
10.0 NTU
Accuracy:
± 2% NTU
Method:
Nephelometric, with white LED and IR LED (860 nm) light sources
Calibration:
Two-point calibration against standard solution
Sensor Body Material:
CPVC
Temperature Limits:
40–120 °F (4–49 °C)
Max. Pressure:
100 PSI (6.9 bar)
Sample Flow
Min. = No minimum
Max. = Maximum carrying capacity of the sample line.