Damaged Circuit Board:

  The circuit boards can be damaged if high voltage is 

connected to the wrong terminals, by lightning, other power surges, or by corrosion.  If 
you believe that the circuit board is damaged, then contact the factory and your local 
sales representative.

 7. 

Damaged Power Supply Board:

  The power supply board accepts AC power (110-240 

V at 50-60 Hz) and provides 24 VDC power to the main circuit board.  If the main circuit 
board does not have power, then check the DC voltage on the output pins of the power 
supply board.  If it has A/C power coming in, but is not putting out 24 VDC, then it either 
has a blown fuse or is damaged and requires replacement.

 8. 

Blown Fuse:

  If the power supply board is not putting out 24 VDC, then always check to 

see if the fuse is blown and replace if necessary.

IV.  OPTIONAL DATA LOGGER

 

1. Description: 

 When enabled in the software, the data logger records the measured NTU of 

each attached turbidity sensor at a selectable frequency.  This data is recorded on the Micro 
SDHC memory card and can be retrieved using any text-reading program.  The Micro SDHC 
memory card is installed in the slot on the MB220 board as indicated on Figure 3 of this 
manual.  To use the data logger the controller must be provided with the MJ500 Real-Time 
Clock board (which mounts directly on the MB220 board as shown on Figure 3).

 

2. Operation:

  To enable, enter the con

fi

 guration menu on the monitor and select the option 

“DL”.  The 

fi

 rst menu option that appears will be the On/O

  menu.  The menus which follow 

allow for adjustment of the data logger frequency and for changes to the clock (date and 
time).  See 

fi

 gure 2.

  

a. 

Frequency

:  The frequency is the time interval between data recordings.  The fre-

quency is adjustable in seconds, with a minimum setting of 5 seconds.

  

b. 

Data Logger Clock:  

The clock is factory-set before shipment.  However, because 

the clock is set on Eastern Standard time it may be necessary to change the date and 
time upon start-up. 

 

3.  Stored Data Files:

  The data will be written to text 

fi

 les on the Micro SDHC memory card.  

The formatting and handling of these 

fi

 les is as described below:

  

a. 

File Format:  

The following is an example data 

fi

 le to illustrate the format used.  As 

you can see, there is a three line header for each 

fi

 le.  The fourth and 

fi

 fth lines are 

headers for the data.  You will see that each header and data entry is delimited by a 
comma.

  

b. 

File Name:  

Each data 

fi

 le will be named according to the date on which it was 

created.  For example if created on May 24, 2022, the 

fi

 le name would be May24_22.txt

 

i.  If the Micro SDHC memory card already has a 

fi

 le started earlier on the same day, 

then data will be written onto the existing 

fi

 le.

 ii. 

The 

text 

fi

 les are limited to 5 MB.  Once this limit has been reached, a new 

fi

 le will 

automatically be created to allow data to continue to be written.

Содержание TH-4000

Страница 1: ...nd Maintenance Manual TH 4000 Rev 4 29 2022 v2 The information contained in this manual was current at the time of printing The most current versions of all Hydro Instruments manuals can be found on our website www hydroinstruments com ...

Страница 2: ... A Entering Setup 5 B Configuring Each Channel 5 IV Maintenance A Sensor Maintenance Precaution 7 B Sensor Cleaning 7 V Troubleshooting A Installation Check 7 B Symptoms Likely Causes and Suggested Responses 7 VI Optional Data Logger 8 Figures 1 Operation Mode Screens 5 2 TH 4000 All Screens 6 3 TH 4000 Controller Electronics 9 4 TH 4000 Circuit Board MB124 10 Drawings TH 4000 Exploded View and Bi...

Страница 3: ...ith a compression fitting designed to ensure correct positioning of the Series sensor in the water stream The sensor has a short channel that can be easily cleaned and calibrated Specifications Specifications TSH 10 TSH 100 TSH 1K TSH 10K Turbidity Range 0 10 NTU 0 100 NTU 0 1 000 NTU 0 10 000 NTU Resolution 0 05 NTU 0 10 NTU 1 00 NTU 10 0 NTU Accuracy 2 NTU Method Nephelometric with white LED and...

Страница 4: ... 7 LED Indicators The red alarm indicator LED on the front panel will illuminate as long as an alarm condition is present B Alarms and Output Signals 1 Acknowledgement of Alarms If an alarm condition occurs the alarm red LED will illuminate and the relay will be activated To acknowledge an alarm and thereby de activate the relay contact output press the key NOTE Even after acknowledging the alarm ...

Страница 5: ...anged using the following procedure A Entering Setup All parameters are set in the password protected setup section 1 Press the key until the password screen is reached 2 Use the and keys to set the password The password is 40 3 Press the key B Configuring Each Channel See Figure 2 The selected channel will flash Move between the channels with the and keys When the desired channel is flashing pres...

Страница 6: ...Filter 30 secs Turb 2 High Alarm 250 NTU Turb 2 Avg Filter 30 secs HOLD HOLD HOLD Modbus Baud 19200 Node 1 Data 8 N 1 Modbus Baud 19200 Node 1 Data 8 N 1 Modbus Baud 19200 Node 1 Data 8 N 1 2400 4800 9600 19200 38400 57600 115200 250000 8 N 1 8 N 2 8 E 1 8 O 1 HOLD HOLD HOLD Turb 1 Zero Cal Skip Hold Begin Turb 1 Zero Cal Skip Hold Begin Turb 1 Span Cal Skip Hold Begin Turb 1 Zero Cal 0 00 NTU Tur...

Страница 7: ...ical sensor channel V TROUBLESHOOTING A Installation Check Review the following points first Sensor Installation Check that the sensor head and flow cell are clean Check that the sensor is properly installed in the flow cell and that water is flowing B Symptoms Likely Causes and Suggested Responses 1 Filter Time If the displayed reading is not stable then the filter time can be increased to elimin...

Страница 8: ...e MB220 board as shown on Figure 3 2 Operation To enable enter the configuration menu on the monitor and select the option DL The first menu option that appears will be the On Off menu The menus which follow allow for adjustment of the data logger frequency and for changes to the clock date and time See figure 2 a Frequency The frequency is the time interval between data recordings The fre quency i...

Страница 9: ...der to eliminate the 3 line file header you can select 4 here to start the data import on row 4 of the file 3 Then click Next v On the next pop up window Text Import Wizard Step 2 of 3 you need to select the type of delimiter being used in the data file The data entries in these files are delimited by commas and so you must select Comma After selecting Comma and only Comma then click Next vi On th...

Страница 10: ... For 120VAC or 240VAC Power Cable Connect at Terminal Block Ribbon Cable from MB220 Power Switch AC Power Terminal Block Cable to Turbidity Probe Micro SDHC Card Slot Ribbon Cable from Membrane Switch MB220 Display Board MJ500 Real Time Clock Modbus Terminals V A B V AC Power Inlet Power Supply Board 24VDC Power Output ...

Страница 11: ...Y B OARD MB124 V V AI2 V AI1 V V AO1 V AO2 CO1 NO1 Relay 1 NC1 Normally Closed CO1 Common NO1 Normally Open NC1 CO2 NO2 Relay 2 NC2 Normally Closed CO2 Common NO2 Normally Open NC2 Analog Output 1 2 AO1 V AO2 V Turbidity Sensor White AI1 Red V Green Black V FIGURE 4 TH 4000 Circuit Board MB124 ...

Страница 12: ...antity No 1 Probe Nut 1 TFC PN 1 2 Turbidity Sensor 1 See Table 3 O Ring 1 TFC BOD OR 4 Flow Tee Body 1 TFC BOD 1 5 O Ring 2 TFC UE OR 6 Flow Tee Union End 2 TFC UE 7 Flow Tee Union Nut 2 TFC UN 8 Tubing Connector 2 10 6 4 Turbidity Sensor No Range TSH 10 0 10 NTU TSH 100 0 100 NTU TSH 1K 0 1 000 NTU TSH 10K 0 10 000 NTU ...

Отзывы: