Recordings
HMG 4000
Edition 2016-03-08 - V01 R04
[59]
Part No.: 669950
8.2.2 Single recording
For a single recording you
must specify the following
settings:
Measurement channel settings
Under this menu point you can set which measurement channels should be added
to the recording and which channels are to be "
switched off
" for it.
Measurement rate
Under this menu point you can set the time interval that the HMG 4000 reads in a
measured value from the "
active
" sensors and stores it in the internal memory.
One recording can comprise a maximum of 8 million measured values.
Selecting 1 ms, for example, means that 1000 measured values per second are taken
and saved for each connected measurement channel.
The measured values according to the set measurement rate are saved
independent from the signal update time of the connected sensors.
This means that, for example, with a preset measurement rate of 1 ms and a
signal update time of 10 ms, the measured value will be saved 10 times before
the signal at the input changes.
Maximum recording duration
Under this menu point, you can select a recording length of 10 seconds, staggered
for up to 28 days.
Zoom area
With the "zoomed area", an additional display area can be created during the
recording in the graphic mode.
This means that the X axis in the "
Zoomed
" setting is scaled to the time range set in
this menu point (switched off / 5 seconds / 10 seconds / ...).
Storage settings
Under this menu point you can specify the storage type, folder and file name for
the corresponding recording.