ALPC 9000 series
Setting up the ALPC
Restrictions pertaining to use
WARNING
Fire hazard
Danger of bodily injury
►
Only test / check fluids with a flash point
55°C /
131°F 56°C / 133°F 55°C / 131°F
NOTICE
Unpermitted fluids
The ALPC will be damaged.
►
The ALPC is only to be used with mineral oils or mineral oil-based
raffinates.
►
Only use liquids which are compatible with the installed materials and the
sealing materials.
►
Only use fluids with a flashpoint 55°C / 131°F
►
Ensure that the flushing fluid has a substantially better cleanliness class
than the cleanliness class expected from the sample (e.g. ISO 15/12/9).
Setting up the ALPC
In order to set up and operate the ALPC, please provide the following at the building
site:
- Supply voltage 230 V AC / 50 Hz
- Clean, dry compressed air
- Space or a table with a load capacity of at least 260 kg
- Flushing fluid tank
- Return
tank
HYDAC FILTER SYSTEMS GMBH
en
Page 16 / 68
BEWA ALPC9000 3601005 en 2011-04-07.doc
2011-04-06