Performing inspection / maintenance
Cleaning the analysis chamber
The cleaning cycles of the analysis chamber are dependent on the test liquid
used.
Used test liquid
Daily
Weekly
Water containing surfactants
X
AIII cleaners or similar, such as G 60 Spezial
X
For cleaning, we recommend a moist, clean, commercial microfiber cloth.
Finish cleaning with a long flushing of the inner chamber.
Clean the diffuser on the filter membrane holder
Clean the filter membrane holder at weekly intervals. If necessary, clean
sooner.
1. Open the coupling (1) and remove the
diffuser (3).
2. Clean the diffuser (3) by blowing it out
with compressed air.
3. Put the diffuser into the filter membrane
holder and screw the coupling (1) finger
tight.
CTU 1x3x Series
en(us)
Page 65 / 92
BeWa CTU1x3x 3631694d en-us 2017-03-13.docx
2017-03-13