16 / 27
Name (number), function
Adjustable value range
Description
Default value
5.000 m (16.400 ft) (level)
HyView setting:
“
Device Settings
” > “Outputs” > “
De-
Energized value”
VGF-DISPLAY:
MAIN MENU / OUTPUT SETUP / RELAY OUTPUT / DEENERGIZED VALUE
RELAY delay (P16)
0…999 sec
In the event that the PV measured value has reached the lower or upper switching value or an error
signal has occurred in the event of an error, the RELAY is activated and a change is visible on the
relay output after this time delay.
HyView setting:
“
Device Settings
” > “Outputs” > “Relay delay time”
VGF-DISPLAY:
MAIN MENU / OUTPUT SETUP / RELAY OUTPUT / DELAY
0 sec
Tank type (P40a),
Tank shapes for volume
measurement
Selectable values:
−
Output Conversion table (OCT)
−
Standing cylindrical tank with dome
bottom
−
Standing cylindrical tank with
conical bottom
−
Standing rectangular tank with or
without chute
−
Lying cylindrical tank
−
Spherical tank
Selecting a typical basic container shape for volume measurement. The tank dimensions can be set
using parameters P41… P45 (see figures below). If OCT is set, the tank shape must be specified in
tabular form.
HyView setting:
“
Device settings
” > “Tank/Silo parameters” > “Tank shape”
VGF-DISPLAY:
MAIN MENU / CALCULATION / TANK SHAPE
IT IS ONLY DISPLAYED IF THE PV IS VOLUME!
Standing cylindrical tank with dome
bottom
Tank type (P40b), Tank
properties for volume
measurement, tank bottom
shape
Selectable values:
−
0 (flat)
−
1
−
2
−
3
Assigning typical tank bottom designs to a specific tank type for accurate volume calculation. The
exact form for the setting code is shown in the figure below.
HyView setting:
“
Device settings
” > “Tank/Silo parameters” > “
Bottom
shape”
VGF-DISPLAY:
no adjustment available
0 (flat)
Standing cylindrical tank with dome
bottom
Standing cylindrical tank with
conical bottom
Standing rectangular tank
Lying cylindrical tank
Spherical tank
Tank dimensions (P41…P45), for
volume measurement
0…999 999
Typical dimensions in units of length for the tank type set in parameter P40 for volume calculation.
HyView setting:
“Device settings” > “Tank/Silo parameters” > ”Bottom shape”
VGF-DISPLAY:
no adjustment available
0
Specific gravity (P32),
for weight measurement
0.
01…100
If the device is set to weight measurement, the specific gravity of the material (medium) stored in the
tank must be entered here for weight calculation. The value to be set is the ratio (without unit) to the
density of the water.
HyView setting:
“
Device settings
” > “Measurement optimalization” > “Specific gravity”
VGF-DISPLAY:
no adjustment available
Threshold offset (P34), interference
sensitivity setting
-
4095…+4095
Modifies the acceptance threshold set on the echo chart. It can be used it to increase (positive value)
or decrease (negative value) the device's ability to suppress the interference signal relative to the
default setting. When set to 0, there is no change from the original setting.
HyView setting:
“
Device settings
” > “Measurement optimalization” > “
Threshold
offset”
VGF-DISPLAY:
MAIN MENU / MEAS. OPTIMIZATION / THRESHOLD OFFSET
0