
8 / 10
110
Read all dynamic variables
8.
TROUBLESHOOTING
Unit is displaying incorrect distance
Confirm display mode is suitable.
Distance
is measured from sensor face to measured level.
Material
is measured from Low-Level to measured
level.
If unit is measuring too deep increase
Sensitivity
. If the unit continues to measure too deep, increase
WaveBoost
.
If unit is measuring too high, reduce
Sensitivity
and ensure
WaveBoost
is set to 0.
Check if material is present to be measured. Unit will output and display FailSafe reading if it cannot detect a level within range.
Analogue trend is erratic/unstable
Increase
Damping
value for smoother trends.
Choose a slower
Tracking
speed.
Check there are no objects interfering with the transit pulse in the application (such as ladders and cross beams).
Unit is locked up or flat line measurement
Confirm there is material within the measurement range, the unit may go to failsafe if there is no flat level available.
If the unit is locked to a high level perform
Mapping
for distance beyond false echoes.
If the unit is locked higher than the real level check for objects interfering with the transit pulse in the application (such as ladders and cross beams).
Lower the
Sensitivity/WaveBoost
. Re-locate the installation to avoid the interfering object.
If the unit is locked lower than the real level confirm the application is not within
Blanking
distance. Increase
Sensitivity/WaveBoost
until the unit
measures correct level.
Tracking
speed may be too slow for the application. Increase the
Tracking
speed.
PLC indication does not match measurement
Connect a Multimeter in series with the powered loop. Compare the 'mA' diagnostic on the display with the mA reading on the loop. If these values do
not match, disconnect the loop wires and measure the resistance across the loop. This should not exceed specification.
Confirm
High-Level
and
Low-Level
are set to the same values in MiniWave and control system.
AutoSet fails
Increase the
Sensitivity
to a higher value.
Allow the unit to run for a longer time (one minute). Re-attempt the
AutoSet
.
Increase
WaveBoost
and re-attempt the
AutoSet
.
Set High- and Low-Level manually in the
Setup
menu.
If problems persist, install a replacement
Microflex-D
in the application, and bench test the problem unit in a controlled environment. Difficult applications may
require the use of different level measurement technology; consult a
Hycontrol
engineer in order to find the most appropriate solution for your requirements.
9.
ACCESSORIES
2” ANSI Polypropylene Flange for NPT threaded units
2” ANSI Polypropylene Flange for BSPT threaded units
Please consult
Hycontrol
for flange order codes.