5.2.2.
Frame Tracing
①
Place the frame between the upper grippers and the
lower grippers.
②
Move the frame in order that the bridge of frame is
located at the central position between the left and
right side grippers.
③
Press the Tracer UI button on the main screen.
④
Input job number and select the frame material.
⑤
Select reading side and input bridge size for one eye
tracing.
⑥
Press the reading button.
5.2.3.
Pattern Tracing
①
Make a marked spot on the optical center of the demo
lens by using Lensmeter.
②
Block the Lens Adaptor at the center of the marked
spot by using Blocker.
③
Place it on the Pattern Holder.
④
Place the pattern holder on the tracer.
⑤
Press the Tracer UI button on the main screen.
⑥
Input job number and select the curve mode.
⑦
Select reading side and input bridge size.
⑧
Press the reading button.
[How to place the Pattern Holder on the Gripper]
•
The data will be automatically registered in the job manager after finishing the tracing.
•
With one side tracing, the data for R or L will be symmetrically copied to the data of the
other side automatically.
Demo Lens
Lens Adaptor
!
NOTE
Содержание HPE-410
Страница 1: ...PATTERNLESS EDGER HPE 410 USER MANUAL...
Страница 145: ...145 HPE 410 8 4 2 X Y R You can check X Y R axis X axis test Y axis test R axis test X Y R sensor state 1 2 3 4...
Страница 146: ...146 8 4 3 Feeler You can check the Feeler Initialize feeler Feeler test Feeler sensor and switch state 1 2...
Страница 149: ...149 HPE 410 8 4 6 Clamp You can check clamp Clamp test Clamp sensor state 1 2...
Страница 166: ...166 10 3 Drawings of System 10 3 1 Drawings of Edger...
Страница 167: ...167 HPE 410 10 3 2 Drawings of Blocker HMB 8000 10 3 3 Drawings of Blocker CBK 4000...