163
HPE-410
Cannot select the
auto bevel position
Do not have the information of frame curve.
Enter the R/L frame curve values at job
editor.
Bevel/groove
position is break.
If the lens has big diameter and high curve, it
can be interference with instruments. so it
cause error when calculate the position.
Apply the pre-roughing mode if occurred
interference when feeling.
The lens stops on
the rouging wheel.
The main contact sensor get the foreign
substance.
Check the main contact sensor at the
test mode.
Incorrect system
time
Reached end of battery life.
(Life time : 5 years. May be reduced
depending on the usage environment)
Replace the GUI board battery.
Type : CR1210