![Huvitz HDR-9000 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/huvitz/hdr-9000/hdr-9000_service-manual_2190280007.webp)
- 6 -
2.2 Setting-Up System Parameters
It is very common to interface HDR-9000 Digital Refractor with several different optometric
instruments. But it varies from customer to customer that what kinds of and how many instruments
are interfaced with. So it’s required for the engineer who installs and sets up refraction system at
customer’s place to set up initial system parameters according to customer ’s needs and environment.
Parameters listed below are the most common parameters required to be set after installation.
Language
Print To Serial (Whether to use PC interaction or not)
Chart notation (decimal / feet / meter) & Controller type
Reset Type On Clear
Chart device
Chart connection type (serial / IR)
[System Setting]
Содержание HDR-9000
Страница 1: ...Service Manual Digital Refractor HDR 9000 HUVITZ Co Ltd...
Страница 2: ...1...
Страница 9: ...8 Engineer Mode Refer to appendix for more detailed information about this mode...
Страница 17: ...16 Unscrew M2 6 screws and remove BACK CAP BACK CAP...
Страница 18: ...17 Unscrew 3 M2 6 screws and remove TOP FRONT CASE TOP FRONT CASE...
Страница 19: ...18 Unscrew 6 M2 6 screws and remove INNER CASE L INNER CASE R INNER CASE L INNER CASE R...
Страница 29: ...28 4 5 Disassembling the Operation Panel Removing procedure for the bottom case of the Operation Panel...
Страница 30: ...29 4 6 Removing the LCD Panel Cover of the Operation Panel...
Страница 32: ...31 4 8 Disassembling the Junction Box...
Страница 35: ...34 5 Trouble shooting 5 1 Wiring Diagram and Board Description 5 1 1 Operation Panel Wiring Diagram...
Страница 37: ...36 5 1 2 Refractor Body Wiring Diagram...
Страница 40: ...39 5 1 3 Junction Box Wiring Diagram...
Страница 42: ...41...
Страница 57: ...56 Both sides of the Refractor Body Disk Selection Information Lens Selection Number Lens Names...