![Huvitz HDR-9000 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/huvitz/hdr-9000/hdr-9000_service-manual_2190280053.webp)
- 52 -
[Confirm Screen]
When you complete inputting five times, it displays the button five times. To complete the
calibration, You have to click the button five times within 20 seconds. Otherwise go back to
the previous screen(Guide mark). Pressing both the five buttons within 20 seconds, it
complete the touch calibration.
Содержание HDR-9000
Страница 1: ...Service Manual Digital Refractor HDR 9000 HUVITZ Co Ltd...
Страница 2: ...1...
Страница 9: ...8 Engineer Mode Refer to appendix for more detailed information about this mode...
Страница 17: ...16 Unscrew M2 6 screws and remove BACK CAP BACK CAP...
Страница 18: ...17 Unscrew 3 M2 6 screws and remove TOP FRONT CASE TOP FRONT CASE...
Страница 19: ...18 Unscrew 6 M2 6 screws and remove INNER CASE L INNER CASE R INNER CASE L INNER CASE R...
Страница 29: ...28 4 5 Disassembling the Operation Panel Removing procedure for the bottom case of the Operation Panel...
Страница 30: ...29 4 6 Removing the LCD Panel Cover of the Operation Panel...
Страница 32: ...31 4 8 Disassembling the Junction Box...
Страница 35: ...34 5 Trouble shooting 5 1 Wiring Diagram and Board Description 5 1 1 Operation Panel Wiring Diagram...
Страница 37: ...36 5 1 2 Refractor Body Wiring Diagram...
Страница 40: ...39 5 1 3 Junction Box Wiring Diagram...
Страница 42: ...41...
Страница 57: ...56 Both sides of the Refractor Body Disk Selection Information Lens Selection Number Lens Names...