Product Specification
FM800Encore Service Manual
8
4 Product Specification
4.1
Physical Specification
Height (all models)
186mm (7.3in)
Length (all models)
358mm (14.1in)
Width (FM820E)
363mm (14.3in)
Width (FM830E)
392mm (15.5in)
Weight (FM820E)
Gross weight 13.5kg (includes transducers etc) Net weight 9.5kg
Weight (FM830E)
Gross weight 15.5kg (includes transducers etc) Net weight 10.5kg
4.2
Environmental
Operating temperature
+10°C to +35°C
Storage/transport temperature
-20°C to +50°C
Operating pressure
68 to 106 kPa (680 to 1060 mB)
Storage/transport pressure
68 to 106 kPa (680 to 1060 mB)
Operating humidity
10% to 75% RH, non-condensing
Storage/transport humidity
10% to 90% RH, non-condensing
4.3
AC supply voltage and fuse values
Rated AC supply voltage
100-120V, 220-240V, 50-60Hz
Fuse values
T2.5A L 250V for nominal input voltage
≥
200V
T2.5A L 250V for nominal input voltage < 200V
Power rating
100VA
4.4
Ultrasound Transducer
Range
30 to 240 bpm
Accuracy
± 1 bpm over the range 100-180 bpm
Alarms
High and Low FHR: 30 to 240 bpm Signal
Loss: % loss in last 5 minutes
Mode
Directional pulsed Doppler Repetition rate 3.0KHz
Frequency
1.0MHz (green)
P-
<1Mpa
Iob
<20mW/cm2
Ispta
<100mW/cm2
Resolution
12 bits
Safety
Type CF protection