UltraScan PRO Supplemental Manual for EasyMatch QC Ver. 2.0
16
Figure 16. Setup Sensor
4.
Remove the tape covering the reflectance port and the exterior of the sphere in the
transmittance compartment.
5.
Place the desired port plate at the reflectance port and snap it into place.
6.
Locate the reflectance port sample clamp and slide it into place at the reflectance port.
Depress the back button to slide the clamp all the way in. Depress the front button to move
the sample clamp up against the sphere. Covering the sample port when the instrument is
not in use prevents dust from accumulating inside the sphere.
7.
Turn on the UltraScan PRO by switching the on/off switch on the back of the sensor to the
on position. Allow the instrument to warm up for two hours prior to standardizing and
making measurements.