Aeros Supplemental Manual for EasyMatch QC ver. 2.0
19
Cleaning the Instrument Standards
It is extremely important that the standard tile (the slide-on calibrator) be treated with great care. It
should always be protected from physical damage and dirt .
Be sure the standards are clean before using it for standardization. Once a month, use a soft nylon
brush and a Sparkleen (Fisher Scientific Catalog Number 4-320-4) in water solution (1 tablespoon per
gallon), followed by a running hot water rinse. Wipe dry with a clean, lint-free paper towel
manufactured without fluorescent whitening agents (FWAs) to remove water marks.
Diagnostics on the Aeros
Diagnostics are included with the Aeros EasyMatch QC. To initiate, select
DIAGNOSTICS
from the
SENSOR MENU.
THE GREEN TILE TEST, REPEATABILITY TEST
and
HARDWARE CHECKS
are the available
tests.
Green Tile Test
To begin the Green Tile Test, standardize the instrument and place the Green Tile at the port. The
software first prompts for the values on the back of the tile.
Figure 18. Enter Green Tile Values
Remove all samples from the instrument and attach the standardization box. The sensor will move to the top.
Содержание Aeros EasyMatch QC