Warnings (continued)

Tipping hazards

Installing any monitor arm con guration not tested for use with the product (i.e. not listed in the product
Speci cation Guide) may result in an unsafe product and increase the likelihood of the cart tipping over. Non-
tested monitor arm con gurations must not be installed.
When using arm-mounted technology, the arm should always be extended to the front of the cart. Extending the
arm to the side or rear of the cart increases the likelihood of the cart tipping over. Speci cally, tipping is likely
under the following conditions:

The work surface platform is in the upper quarter of its height adjustment range.
A double arm con guration is extended out from the rear of the cart.
The rear casters are locked.
Weight is biased towards the rear of the cart.
The user applies a forwards push force (from the front of the cart) that is more than 2x the typical push
force.

When connecting power cables from technology on the cart to wall AC outlets or similar the cart must remain
static and not be moved around. If the cart is moved away from the wall outlet while the power cable is still
connected the power cable may reach its maximum extension and cause the cart to become unstable and tip over.
The cable storage basket, its contents, any attached cart accessories and the technology mounted to the cart may
also be damaged or dislodged in this scenario.
The cart may become unstable and increase the likelihood of tipping over when pushed over thresholds or changes
in  ooring with a height difference of 1.5" (40mm) or greater.
The work surface should not be loaded beyond the weight limit listed in the product technical information. It
should not be used to support a person's body weight or as a seat.
Failing to adhere to any of the above guidance means the cart is more likely to tip over and cause physical injury or
damage to property.

Warnings continued on next page.

13

Back to Contents

Содержание TouchPoint T4

Страница 1: ...SCALE HEALTHCARE CUSTOMER SUPPORT Americas 1 800 400 0625 Option 4 hcorders humanscale com International 1 303 858 9912 Option 4 hcorders humanscale com humanscalehealthcare com h s h c t 4 h a r d w a r e i n s t a l l a t i o n g u i d e e n u s 2 0 2 0 0 8 1 7 ...

Страница 2: ...400 0625 Option 4 hcorders humanscale com International 1 303 858 9912 Option 4 intlhcorders humanscale com Visit the Humanscale Healthcare website for Terms and Conditions https www humanscale com about legal information terms conditions healthcare cfm 2 Contents 2 Components List 3 Assembly Instructions 5 Cleaning Instructions 11 Warnings 12 ...

Страница 3: ...p Lock Bracket Post for Ball Joint Mount Ball Joint Assembly VESA Plate Assembly Post Mount Post for Arm Mount Monitor Arm Bracket Monitor Arm Straight Link Number of parts received varies based on con guration Monitor Arm Dynamic Link Tilt Mechanism is included 3 Back to Contents ...

Страница 4: ...COMPONENTS CONTINUED Supplied components may vary depending on product con guration Monitor Arm Tilt Mechanism VESA Plate Assembly Arm Mount Cable Management Clip 2 pieces 4 Back to Contents ...

Страница 5: ...lide between the underside of the work surface and the accessory rail STEP 3 Secure the Laptop Lock Bracket to the underside of the work surface on both sides using the two size 10 x 0 75 self tapping Phillips screws provided NOTE There are 3 sets of pre drilled holes where the screws can be installed Use the set of pre drilled holes that is appropriate for the size of your laptop This determines ...

Страница 6: ...ide the Cable Management Clips over the Post and squeeze them to lock at the location of your choice CAUTION It is important to use the Lock Washer and place it between the bottom of the Post and the work surface STEP 2 Attach the VESA Plate Post Mount to the rear of your technology device e g Tablet Holder or similar using the four M4 Phillips screws 6 Back to Contents ...

Страница 7: ...he VESA plate into the Ball Joint Assembly until it snaps into place To release pull the plastic tab back on the Ball Joint Assembly and lift the VESA plate and technology device upwards Tighten the two M6 hex set screws at the rear of the Post to secure the Ball Joint Assembly in place and prevent it from rotating 7 Back to Contents ...

Страница 8: ...use the Lock Washer and place it between the bottom of the Post and the work surface STEP 2 Open the lever on the side of the Monitor Arm Bracket Slide it over the Pole to the location of your choice Close the lever to secure the bracket on the Post Adjust the Smart Stop Rotation Ring to set rotational stop if required To select the appropriate orientation of the ring read the rotation angles mark...

Страница 9: ...s as detailed in Step 2 above Adjust as required to set the appropriate rotational stop angle STEP 4 Remove the Plastic Cover from the VESA Assembly Arm Mount Mount the VESA Plate on the technology device e g Tablet Holder or similar using the four M4 Phillips screws Snap the plastic cover back onto the VESA Plate STEP 5 Locate the VESA Assembly onto the feature at the top of the Tilt Mechanism Al...

Страница 10: ...ce up and down STEP 2 If the work surface moves up easily but is hard to push down Using the provided hex key 1 turn the adjustment screw 2 counter clockwise to move the indicator 3 towards Minimum until the work surface is stable and requires the same force to move up and down STEP 3 If the work surface is hard to pull up but drops down easily Using the provided hex key 1 turn the adjustment scre...

Страница 11: ...ution Dextrose 5 Glucose solution HCl 0 5 solution pH 1 Hydrogen Peroxide 6 solution NaCl Solution 0 9 solution Electrode Gel 98 water 2 Carbopol 940 solution Warnings Always unplug the cart and power down technology devices before cleaning Never clean the cart with steel wool or other abrasive materials that will damage the surface nish Do not rub a wet or damp cloth over any electrical component...

Страница 12: ...ted technology e g a tablet cables should be secured in the cable management clips attached to the post When routing cables to technology on the work surface e g a laptop care should be taken to avoid trapping cables in and around the laptop locking bar Failure to adhere to this guidance may lead to cables or device chargers impacting or snagging on people or items in the use area leading to damag...

Страница 13: ...s more than 2x the typical push force When connecting power cables from technology on the cart to wall AC outlets or similar the cart must remain static and not be moved around If the cart is moved away from the wall outlet while the power cable is still connected the power cable may reach its maximum extension and cause the cart to become unstable and tip over The cable storage basket its content...

Страница 14: ... when moving or adjusting the height of the cart in close proximity to patients Take care when moving the care especially when pushing the cart forward that the cart or any mounted technology does not obscure the view in front of the cart to avoid colliding with people or objects in the path of travel If the cart collides or impacts with another hard object it may damage the cart in particular the...

Отзывы: