Interfaces and software update
Chapter 6
OPERATION MANUAL
V2.4.0en/22.07.20//17.12
Unistat® T300 Series
83
6.3.1
Jack LEVEL (Com.G@te® external only)
For level monitoring in the
>Sight glass<
[23].
This connection enables you to connect an external float switch (Order No. 6152), which is posi-
tioned in the
>Sight glass<
[23], for monitoring the level of your externally closed application. Acti-
vation via a potential-free contact.
The interface is specified as a digital input. Do not apply voltage or current.
Pin
Description
1
Level test (bypass via contact 2 → “absence”)
2
Level – (GND)
3
Level + (normally open contact)
6.3.2
Connector POKO (floating contact) alarm
Information on the interface can be found on page 79 in Section
»Connector POKO (floating con-
.
6.3.3
Jack AIF Reg-E-Prog
The analog interface has a programmable input channel and 3 output channels.
The analog interface of Com.G@te® is programmed in the category “Interfaces”.
Pin
Description
Signal
1
Current output, T external
0
a)
/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω resistor.
2
Current output, setpoint
0
a)
/4 - 20 mA or 0 - 10 V
3
GND for analog outputs
GND
4
Analog input (programmable)
0
a)
/4 - 20 mA or 0 - 10 V
Power input: 200 Ω working
resistance
Voltage input: 100 kΩ input
resistance
5
Current output, freely programmable
0
a)
/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω resistor.
6
GND for analog input
GND
a)
Consult with our Customer Support.
Pin assignment (front
view)
Pin assignment
Pin assignment (front
view)
Pin assignment
Содержание Pilot ONE Unistat T3*0
Страница 1: ...Unistat T300 Series...
Страница 2: ......
Страница 3: ...OPERATION MANUAL Unistat T300 Series...
Страница 4: ......
Страница 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Страница 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...