Setup mode
Chapter 4
OPERATION MANUAL
V2.4.0en/09.08.21//1.30
Minichiller®
37
4
Setup mode
4.1
Setup mode
Moving the temperature control unit during operation
SERIOUS BURNS/FREEZING OF THE HOUSING PARTS/ESCAPING THERMAL FLUID
Do not move temperature control units that are in operation.
4.1.1
Turning on the temperature control unit
PROCEDURE
Switch on the temperature control unit using the
>Mains switch<
[37].
Circulation and thermoregulation are turned off.
4.1.2
Turning off the temperature control unit
PROCEDURE
Warm the thermofluid to room temperature.
Stop the thermoregulation.
Switch off the temperature control unit using the
>Mains switch<
[37].
4.1.3
Setting the overtemperature (OT) protection
The overtemperature protection is set higher than the ignition temperature of the thermal fluid
used
MORTAL DANGER FROM FIRE
The overtemperature protection must be correctly set to the thermal fluid you are using.
Always observe the safety data sheet of the thermal fluid.
Set the cut-out value of the overtemperature protection at least 25 K below the fire point of
the thermal fluid.
4.1.3.1
General information on the overtemperature protection
The overtemperature protection is installed only in temperature control units that have a heater.
The flow temperature is monitored to ensure the safety of your system. It is set immediately after
you have filled the system with thermal fluid.
Upon delivery, the cut-
out value of the overtemperature protection is set to 40 °C. An alarm is trig-
gered by the temperature control unit shortly after turning on the power if the temperature of the
thermal fluid just filled is higher than the cut-out value set for the overtemperature protection. Set
the overtemperature protection to the thermal fluid you are using. Please note: The printed scale
can deviate by - 25 K from the set cut-out value.
Example of a
potentiometer at the
temperature control
unit