Commissioning
Chapter 2
OPERATION MANUAL
V2.4.0en/09.08.21//1.30
Minichiller®
25
Use of the temperature control unit is permitted only under normal ambient conditions in accord-
ance with the currently valid DIN EN 61010-1.
▪
Use only indoors. The illuminance must be at least 300 lx.
▪
Installation altitude up to 2,000 meters above sea level.
▪
Maintain wall and ceiling clearance for adequate air exchange (dissipation of waste heat, supply
of fresh air for the temperature control unit and work area). Ensure adequate floor clearance for
air-cooled temperature control units. Do not operate this temperature control unit from within
the box or with an inadequately dimensioned bath. This inhibits the air exchange.
▪
Ambient temperature values are provided on the technical data sheet; to ensure trouble-free
operation, compliance with the ambient conditions is mandatory.
▪
Rel
ative humidity max 80% to 32 °C and 40 °C decreasing linearly to 50%.
▪
Short distance to supply connections.
▪
The temperature control unit must not be installed so as to hinder or even prevent access to the
disconnecting device (to the power supply).
▪
For the magnitude of the mains voltage fluctuations, refer to the datasheet.
→
From page 59,
section
▪
Transient surges, as would normally occur in the power supply system.
▪
Installation Class 3
▪
Applicable degree of soiling: 2.
▪
Surge category II.
Please note:
→
»Exemplary illustrations of the cooling variants«
Side
Distance in cm
Air cooling
Water cooling
[A1]
Top
Air outlet on top of unit: free standing
–
[A2]
Top
can be located under a bench
can be located under a bench
[B]
Left
min. 20
min. 10
[C]
Right
min. 20
min. 10
[D]
Front
min. 20
min. 10
[E]
Rear
min. 20
min. 20
Wall clearances