4
Figure 2
[heat capacity]: click the display box behind the heat capacity to pop up the keyboard and
input the heat capacity
The heat capacity of the system is positively proportional to the measured calorific
value. Input the average value of the calibrated heat capacity, click return, save and
return to the main interface of the system.
[standard calorific value]: click the display box behind the standard calorific value to pop
up the keyboard and input the standard calorific value.
The calorific value of the reference material used in the calibration of the instrument is
directly input according to the data on the package of the reference material used. Click
return to save and return to the main interface of the system.
[water injection time]: click the display box behind the water injection time to pop up the
keyboard and input the water injection time.
The water injection time of the inner cylinder is related to the experimental accuracy,
and it is generally set as submerged oxygen bomb. According to the volume and voltage
of oxygen bomb, input the appropriate value, click return, save and return to the upper
menu
[drainage time]: click the display box behind the drainage time to pop up the keyboard
and input the drainage time.
Set appropriate drainage time to drain water for inner cylinder test.