Click the calibration input as follows:
Input the sample weight, number and other test data, and click return to return to the heat
capacity measurement interface.
(4). [return]
When the test is over and ignition fails, press this key to clear the test information and
curve and return to the standby state.
Note: after using the data processing, click this key to clear the processing trace, so as
not to affect the print results.
(5). [reset]
Press this key and the instrument will return to the power on state.
(6). Thermal capacity calibration
Weigh benzoic acid (0.9 ~ 1.1g), load oxygen bomb, charge oxygen for 15 ~ 45 seconds,
pressure is 2.8 ~ 3.2Mpa, put it into the inner cylinder, click "thermal capacity
measurement", input the weight of benzoic acid, more than ten minutes later, the
instrument will print out the calibrated heat capacity e, repeatedly make 5 ~ 6 pieces,
calculate its average value, input and save in the system, The calorific value of 1 ~ 2