![Huazheng HZ-384A Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/huazheng/hz-384a/hz-384a_instruction-manual_2179912005.webp)
2
Pressure resistance: water pressure 20MPa
Weight: 2.5 kg
Overall dimension: 650 * 450 * 500 mm
Temperature range: 0 ~ 40
℃
Response time: < 4 S
Resolution: 0.0001
℃
Linearity: < 0.08% per 5
℃
temperature rise range
Temperature measurement error: ± 0.003
℃
for every 5
℃
temperature rise range
Power supply voltage: AC220V ± 10%
Humidity: ≤ 80%
Power: 30W
Ignition voltage: AC24 V
Ignition time: 5S
V.Instrument structure and auxiliary equipment
1.Main structure of calorimeter
The host of calorimeter includes: shell, water tank, control system, user interface and so
on
2.The water tank is made of stainless steel and plastic.
VI.How to use the instrument
1.Preparation before use
Open box inspection: after unpacking, check whether the appearance of the instrument is
damaged and whether the accessories are complete. If there are any problems
mentioned above, please contact the manufacturer in time.The calorimeter should be
installed in the room without strong air convection, direct sunlight, and temperature
change should not exceed 1
℃
. The room temperature should be within 15 ~ 30
℃
.
Place the calorimeter on the horizontal ground or on the workbench, and connect the
power supply. Add distilled water with water temperature equal to room temperature until