8
CS2674AX hipot tester is without over zero start function, in the moment of start button
pressed, high voltage output may from the sine wave peak or trough of the sine wave;
which may generate high voltage overshoot. Also, if turn off the power switch, high voltage
overshoot also may be caused. Therefore, using before the tester output the high voltage,
operator should adjust the voltage knob to zero firstly and then start the tester, rotate of
the voltage regulating knob slowly to the required voltage value.
III.Front Panel
(一)
Front panel illustration
(二)
Rear panel illustration
(一)
Front panel
1
、
Power switch: It’s used to control whether turn on/off the power
2
、
Start
:
Pressing it, the test light illuminates and the instrument begins to work.
3
、
Reset: Pressing it, the test light goes out. No high-voltage is output at this moment.
4
、
Remote control interface: Connecting with remote control test switch.
5
、
Voltage knob
:
Rotating it can set the output voltage from 0k V to 20kV
;
Anti-clockwise for small and vice versa for large.