T2000-T2100
EN – 15
4.5.4. HOLD
Shortly pressing the multifunction key “
3
”
activates the “
HOLD
” function and freezes the
result on the display (see Fig. 14). To go back to
the normal measuring mode, shortly press the
multifunction key “
3
” again or shortly press the
multifunction key “
1
” (
ESC
function).
Fig. 14
4.5.5. MEM
Shortly pressing the multifunction key “
2
”
activates the “
MEM
” function, and the result on
the display is saved in the internal memory. For
a few seconds, the screen in Fig. 15 is shown,
which contains the “MEM” symbol and the
number of the location in which this datum has
been saved.
For managing data in the memory, see § 4.8.
Fig. 15
4.5.6. Anomalous situations
While measuring, the indication “
OL
” means
that the resistance measured exceeds the
maximum value which can be measured by the
instrument (see Fig. 16).
Fig. 16
While measuring, the symbol “
” means
that the resistance measured is lower than the
minimum limit for which the accuracy stated in
the manual is granted (see Fig. 17).
Fig. 17
While measuring, the symbol “
” means that
the resistance measured exceeds the maximum
limit set (see Fig. 18).
For managing the alarm thresholds, see § 4.9.
Fig. 18
While measuring, the symbol “
” means
that the instrument has detected a disturbance
current on the resistance measurement loop.
Fig. 19