4.9.
SETTING OF ALARM THRESHOLDS
1. Press and hold the multifunction key “
1
” for more than 2 seconds to switch on the
instrument.
2. Press and hold the multifunction key “
3
” for more than 2 seconds (
SET LIM.
function) to
enter the alarm threshold setting section.
Depending on whether the instrument is set to resistance or current measurement, one
of the following screens will be displayed.
(T2000 only)
Fig. 39: Setting of alarm thresholds for resistance and current measurement
3. Shortly press the multifunction keys “
2
” (
function), or “
3
” (
function), to respectively
increase or decrease the limit value of the alarm threshold.
The limit values which can be set are:
Resistance: from
1
Ω
to
999
Ω
Current: from
1mA
to
999mA
(T2000)
4. Shortly press the multifunction key “
1
” (
ESC
function) to confirm the alarm threshold
value set and go back to measuring mode.