HT9022
EN - 37
5.6. POSITION
“W ”: LOG, ON-LINE SCOPE, SNAPSHOT, MEMORY, DOWNLOAD
WARNING
The maximum DC or AC+DC input voltage is 1000V and the maximum
measurable DC or AC+DC current is 1000A. Do not measure voltages and
currents exceeding the limits given in this manual. Exceeding these limits
could result in electrical shocks to the user and damage to the instrument.
We recommend holding the clamp respecting the safety area created by the
hand protection (see Fig. 3).
Turn the rotary switch to “
W
”. Insert the red cable into the input lead
V
Ω
and the black
cable into the input lead
COM
, and connect the instrument as described in Fig. 12
WARNING
Insert the cable in the middle of the clamp jaws, in order to obtain accurate
measures.
Use the marks as a reference (see Fig. 2).
5.6.1. Key
F3
“Sys”
While measuring
DC
,
AC 1P
or
AC 3P
Power, press key
F3
(
Sys
) to open the drop-down menu shown on the screen to the
side. At each subsequent pressure of key
F3
, the cursor will
scroll through the available items, as follows:
•
Start Log
: it starts a recording of the electrical mains
parameters;
•
On-line
: it starts a Bluetooth On-Line Scope;
•
Memory
: it shows the list of the saved data;
•
SnapShot
: it carries out an instant saving of the measured
Mod Par
Sys
OK
A C 5 0 . 0 H z
2 1 . 4 7
k W
7 . 6 8
k V a r i
2 2 . 9 0
k V A
19/01-17:00:00
parameters;
•
Download
: it sets to the mode for downloading the data saved in the memory;
•
Esc
: it closes the drop-down menu.
By pressing key
F4
(
OK
), the selected item is confirmed.
The electrical parameters recorded during a Recording, transmitted during an On-Line
Scope or saved in a SnapShot, according to the mode set, are the following:
•
AC 1P
:
AC 1P
: P, Q, S, pF, dPf, V, I, THDV, THDI, hV01..hVxx (xx=25 for fundamental
frequency 10..75Hz; xx=8 for fundamental frequency 75..400Hz);
•
AC 3P
: P, Q, S, pF, dPf, V, I, THDV, THDI, hV01..hVxx (xx=25 for fundamental
frequency 10..75Hz; xx=8 for fundamental frequency 75..400Hz);
•
DC
: P, V, I.
5.6.1.1. “Start Log” recording
Upon confirming the “
Start Log
” item, the instrument sets to
stand-by and waits for a recording to start. A recording shall be
started when reaching the following minute as indicated by the
instrument’s time.
The message “
Wait
” is displayed.
Mod Par Sys
Fnz
A C 5 0 . 0 H z
2 1 . 4 7
k W
7 . 6 8
k V a r i
2 2 . 9 0
k V A
W a i t
1 P
19/01-17:00:35
Start Log
Online
Memory
SnapShot
Download
Esc
www.
.com
1.800.561.8187