HT9022
EN - 20
WARNING
If the frequency of the measured voltage is lower than 40Hz or higher than
70Hz, the display shows the message “
F<40 Hz
” or “
F>70 Hz
” and phase
detection does not start.
3. When a voltage higher than or equal to 100V is detected, the
instrument gives an acoustic signal (buzzer) and the
message “
Meas
” is shown on the display. Do not press any
key and keep the test lead connected to L1 phase cable.
Mod Go
P h S e q
P H 1
M e a s
19/01-17:00:00
4. Once phase L1 acquisition is complete, the displays shows
the screen here to the side. Disconnect the test lead from
phase L1 cable.
Mod Go
P h S e q
D i s c o n .
W a i t
19/01-17:00:00
5. The instrument shows the screen here to the side, and waits
for the detection of phase L2.
Connect the test lead to phase L2 cable.
Mod Go
P h S e q
P H 2
W a i t
19/01-17:00:00
WARNING
If more than 3 seconds elapse before detecting phase L2, the instrument
displays the message “
Time Out
”.
It is necessary to repeat the measuring cycle from the beginning, by
pressing key
F4
(
Go
) and starting from step 1 again.
6. When a voltage higher than or equal to 100V is detected, the
instrument gives an acoustic signal (buzzer) and the
message “
Meas
” is shown on the display. Do not press any
key and keep the test lead connected to L2 phase cable.
Mod Go
P h S e q
P H 2
M e a s
19/01-17:00:00