FULLTEST
3
EN - 11
5 ON/OFF mains switch (with red pilot lamp).
6 Connector for START/SAVE remote control adapter.
7 Connector SAFETY INPUT for connection of external safety switch (for example protection door). It
disables DIELECTRIC tests in case the switch is open.
8 IEC female connector for connection of red WARNING LAMP in DIELECTRIC test. The lamp is
active when the test is on (parallel operation to HV lamp on front panel, reference 35).
9 USB 1 connector for connection with PC.
10 USB 2 and USB 3 connectors for connection of USB stick, USB barcode reader, USB printer, USB
keyboard or IMP57 impedance meter.
11 Negative
R
ISO
terminal.
12 Positive
R
ISO
terminal.
13 R
PE
current generator terminal.
14 SENSE voltage terminal.
15 Fuse F4, type F20A/500V 6.3×32 mm which protects internal circuitry in R
PE
measurement.
16 SENSE voltage terminal.
17 R
PE
current generator terminal.
18 FUNC hard key to select appropriate measurement function.
19 SAVE hard key to save test result.
20 START/STOP button which starts or stops selected measurement.
21 RCL hard key to recall saved result.
22 MENU hard key to open MAIN MENU.
23 EXIT hard key to exit existing screen and return it one step back.
24 Colour LCD touch-screen display.
25 Mains test socket for POWER and LEAKAGE measurement.
26 CLAMP connector for HT96 current clamp.
27 U
RES
measurement terminal.
28 U
RES
measurement terminal.
29 L/TRIG/L1 terminal for LOOP, R
A
, RCD, PHASE SEQUENCE and U
RES
measurements.
30 PE/L3 terminal for LOOP, R
A
, RCD and PHASE SEQUENCE measurements.
31 N/TRIG/L2 terminal for LOOP, RCD, PHASE SEQUENCE and U
RES
measurements.
32 DIELECTRIC test terminal for test voltages 2.51
5.10 kV.
33 DIELECTRIC test terminal for test voltages 0.81
2.50 kV.
34 DIELECTRIC test terminal for test voltages 0.25
0.80 kV.
35 DIELECTRIC on lamp. It lights when DIELECTRIC test is running.
36
DIELECTRIC COM (common) test terminal.
4.2. TURNING ON THE TESTER
After switching on mains switch (5) the tester will upload WINDOWS first (it will take 30 s approx.), then
last used measurement screen will appear. A beep-beep sound signal will be given as soon as the
tester is ready for measurements.
CAUTION
In case energizing socket is not earthed properly PE DISCONNECTED message
will appear and the tester will not perform any further operation. In this case
switch off the tester immediately and check the energizing socket!