Test result will be displayed in green colour equipped with green
symbol and with beep-beep
sound if it is lower than or equal to 60 VRMS (may be AC or DC, see the unit). If the result is higher
than 60 VRMS, then it will be displayed in red colour equipped with red
symbol and with longer
beep sound. See the display outlook with test result on the figure 32.
9)
Save the test result by pressing the
SAVE
hard key twice, for further instructions see the
“MEMORIZING EXAMPLE” section.
10)
Press the EXIT hard key to clear displayed result, actual input voltage U
IN
and actual trigger
voltage U
TRIG
will start to be displayed again.
PLUG/1s mode is selected:
4)
Select measurement screen by pressing the
(3) touch-screen key and check all settings again.
5)
Connect test leads according to the figure below.
t
1 s
RES
<
Figure 36: Connection of test leads in U
RES
PLUG measurement
6)
READY, DISCONNECT UUT message will appear when input voltage U
IN
within required range
100
460 V AC is present. Carry out the measurement by disconnecting UUT. Disconnection of
UUT according to the figure 35 means disconnection of multiple plug adapter.
7)
Test result will be displayed in green colour equipped with green
symbol and with beep-beep
sound if it is lower than or equal to 60 VRMS (may be AC or DC, see the unit). If the result is higher
than 60 VRMS, then it will be displayed in red colour equipped with red
symbol and with longer
beep sound. See the display outlook with test result on the figure 32.
8)
Save the test result by pressing the
SAVE
hard key twice, for further instructions see the
“MEMORIZING EXAMPLE” section.
9)
Press the EXIT hard key to clear displayed result, actual input voltage U
IN
will start to be displayed
again.
CAUTION
Do not use the START button in this function, it has
no function!