NEPTUNE
EN - 15
5.5. PHASE SEQUENCE AND PHASE CONCORDANCE WITH 1 TERMINAL
CAUTION
Input AC voltage to carry out this test must be in range
100V ÷ 1000V
with
a frequency in range
42.5Hz ÷ 69Hz
.
This test can only be performed by
touching the metal parts of the
conductors
.
Fig. 10: Use of the instrument for phase sequence and phase concordance test
1. Select position . Message
“
PH 1
”
flashes on the display.
2. Insert the red lead into input terminal
V
/+
.
3. Position the red lead onto phase
L1
of the three-phase system to be tested (see Fig.
10). The following messages may appear on the display (see Fig. 11) to identify the
presence of a voltage signal with frequency out of range
42.5Hz ÷ 69Hz
. In these
conditions, the instrument does not perform the test.
Fig. 11: Signal of voltage with wrong frequency
Содержание F3000U
Страница 1: ... Copyright HT ITALIA 2018 Release EN 1 00 31 10 2018 ENGLISH User manual ...
Страница 2: ......