1. SAFETY INFORMATION

 

 

  

Warning

 

 

Using this instrument should special attention to improper use may cause 

 

electric 

shock or damage, in use should follow the usual safety rules and in full 

 

compliance with the security measures provided for manual. In order to take full   
  

advantage of instrument function and ensure the safe operation, please read these  

  

instructions and follow them strictly   

  
  

1.1 Preparation  
1.1.1

 

Using the instrument the user must comply with standards of safety rules   --

Electric shock protection --to avoid the misuse of instruments   

1.1.2

 

When you received the instrument, check whether the damage in transit   

1.1.3

 

Test lead and test probe must be in good condition connect the test lead before you 

connect the test probe,   when you disconnect them ,disconnect the test probe first   

1.1.4

 

When servicing the meter use only specified replacement parts   

1.1.5

 

Inspect the test lead for damaged insulation or exposed metal, check the test lead 

and test probe for continuity. Replace damaged test lead before you use meter   

 

 

2. FRONT PANEL DESRIPTION  

(1)

 

“LIGHT” button  

(2)

 

“SELECT” button (3)

  

Display 3 1/2 

digits LCD       with max 
reading 1999  

(4)

 

“HOLD” button  

(5)

 

Function switch  

(6)

 

Scalable probe block  

(7)

 

Test probe   

(8)

 

Test lead   

 

 

3. SPECIFICATIONS  

Accuracy is specified for a period of one year after calibration and at 23±5

 with relative 

humidity up to 75% 3.1 Review   

Auto-Ranging  
The max voltage the meter can measure is 600V DC or AC between COM and Earth  
Ground   
Working height: Max 2000m    
Display: 3 1/2 –digit LCD   
Max reading of 1999  
Polarity: Auto polarity indication “-”expressed as negative polarity  
Over range indication: ”OL” on display  
Sampling Rate: Approximate 3 times per sec  Functional 

power unit display   

Low battery indication “

” on LCD  

Fuse protection: mA: Resettable Fuse   

Auto shutdown: 
15minutes   
Battery: 1.5V2LR44  

Operation Temperature 

-

40℃<80%RH

  

temperature: 

-10℃

-

50℃<70%RH

 Remove the battery   

Storage 

■20mA

 - 

0.01mA  

±(1.5%+10d) 

 

■200mA

 - 

0.01mA  

±(1.5%+10d) 

 

■20mA 

  ■200mA

 

 

 

 

 

0.01mA 

 

0.01mA 

 

±(3.0%+10d) 

 

±(3.0%+10d) 

 

AC/DC Voltage 

 

Range  

 

 

Resolution

 

 

   Accuracy

 

 

■200m V  

  

0.1mV  

±(0.8%+5d) 

 

■2 V 0.001V  

±(0.8%+5d) 

 

■20 V 0.01V  

±(0.8%+5d) 

 

■200 V  

0.1V  

±(0.8%+5d) 

 

■600 V  

1V  

±(0.8%+5d) 

 

■2V

 

 

 

■20V 

 ■200V

 

 

■600

 

 

 

 

 

Resistance 

 

0.001V 

 

0.01V 

 

0.1V 

 

1V 

 

±(1.0%+3d) 

 

±(1.0%+3d) 

 

±(1.0%+3d) 

 

±(1.2%+3d) 

 

Range 

 

■200Ω 

 

■2KΩ 

 

■20KΩ 

 

■200 KΩ 

 

■2MΩ 

 

■20M Ω 

 

Resolution  

 

0.1Ω 

 

1Ω 

 

0.01KΩ 

 

0.1KΩ 

 

1KΩ 

 

0.01MΩ 

 

Accuracy

 

 

±(1.0%+5d) 

 

±(1.0%+3d) 

 

±(1.0%+3d) 

 

±(1.0%+3d) 

 

±(1.0%+3d) 

 

±(1.2%+5d) 

 

  

Continuity Test Diode 

test   

3.2 Use   
3.2.1

 

Use the proper function and range for your measurements  

3.2.2

 

When serving the meter, use the same model number or identical electrical 

specification replacement parts  

3.2.3

 

When using the probe s keep your fingers behind the finger guards on the unit   

3.2.4

 

Do not input higher than 600V between the two meter’s input terminal to avoid 

electric shock and damage to the meter   

3.2.5

 

When the meter working at an effective voltage over 60V DC or 30V AC  

(RMS),special care should be taken for there is danger of electric shock  

3.2.6

 

Disconnect circuit power and discharge all high-voltage capacitors before testing 

resistance, continuity and diode  

3.2.7

 

Do not testing resistance, continuity and diode with electricity  

3.2.8

 

Before pushing the function switch to charge functions disconnect test lead and 

test probe from the circuit under test   

3.2.9

 

Do not operate the meter around explosive gas .vapor or dust   

3.2.10

 

Do not use the meter if it operates abnormally ,Protection may be impaired, when in 

doubt have the meter serviced   

3.2.11

 

Do not use the meter with the battery door or portions of the cover removed or 

loosened  

3.2.12

 

Do not store the meter in direct sunlight, high temperature and high humidity  

 

 

4. OPERATION INSTRUCTION  
4.1

 

Reading hold 

 

 

After pressing “HOLD” button

the present reading will be held on the display ,to exit 

from the mode ,just press button again  

4.2 "SELECT" Button

 

 

This button is used for the function's change between: VDC/AC, 

resistance/diode/continuity.

  

4.3

 

Auto Power-off

 

 

If the instrument is not used or stays in a range position for more than 15 minutes, the 

power supply will be switched off and the instrument will be in sleep status. To arouse 
the instrument from sleep, turn the function switch or press the "HOLD" button.  

4.4

 

Measuring Voltage

 

 

4.4.1 Put away the test probe protective cover, and takes test lead from the back 

box.  

4.4.2 Set the function switch in V position  

4.4.3 Press the "SELECT" button to set the unit in DC or AC range.  
4.4.4 Connect the test probe and test lead across the source or load to be measured.  
4.4.5 The display will show the reading along with the polarity of the test probe.  

 

4.5.4 Connect the test lead and test probe across the resistance to be measured.  
4.5.5 Read the value on LCD.  

  

Note: - If the resistance is 2 1M, it takes several seconds to stabilize. It is normal.  
- If the input terminal is in open circuit, overload mark will be displayed on LCD. 

- Before measuring resistance, make sure that the power supply has been switched 

off and all the capacitors have been discharged.

  

4.6 Measuring Diode  

4.6.1 Put away the test probe protective cover.  
4.6.2 Take test lead from the back box.  
4.6.3 Set the function switch toposition.  
4.6.4 Press "SELECT button to make the display show ~  
4.6.5 Connect the test probe and test lead across the diode (test probe to the      

positive pole of the diode, test lead to negative pole).  

4.6.6 Read the forward voltage on LCD. Note:  
- Meter shows an approximation of diode forward voltage drop - If the 

instrument and test lead reverse connection, the LCD will show "OL". - If 
the instrument and test lead open circuit, the LCD will show "OL".  

4.7 Audible Continuity  

4.7.1 Put away the test probe protective cover.  
4.7.2 Take test lead from the back box.  
4.7.3 Set the function switch to '' Ω’ position.  

4.7.4 Press "SELECT button to make the display show “ 

”  

4.7.5 Connect the test lead and test probe across the resistance to be      

measured.  

4.7.6 The buzzer sounds if the resistance of a circuit under test is less than      

50Ω; it may or may not sounds if the resistance of a circuit under test is      
between 50Ω to 120Ω; it does not sound if the resistance of a circuit      under 
test is higher than 120.  

Note:  
- If the test lead reverse connection or the resistance is higher than 200Ω, the LCD will 

show "OL". 

4.8 Current measurement 

 

  

Warning 

 

Risk of electric shock.  

 

When current measurement has been completed, disconnect the connection 

 

between the conductor under test and the jaw, and remove the conductor away 

 

from 

the transformer jaw of the Meter.  
  
  

4.8.1 Put away the test probe protective cover. And take test lead from the back box.  
4.8.2 Set the function switch in mA position  
4.8.3 Press the "SELECT button to set the unit in DC or AC range.  
4.8.4 Connect the test probe and test lead across the source or load to be measured. 4.8.5 

The display will show the reading along with the polarity of the test probe.  

5. MAINTENANCE  

  

  

               

  

Warning  

Dimensions: 205mm×35mm×30mm  
Weight: about 96g (without battery)   
  

AC/DC Current  

 

 

Range

  

Resolution

 

 

  Accuracy

 

 

  

  

  

 

 

 

 

 Measuring Resistance 

 

4.5

 

  

  

  

  

  

4.5.1  Put away the test probe protective cover.  
4.5.2  Take test lead from the back box.  
4.5.3  Set the function switch in the  

position.  

W arning 

Risk of electric shock.  
The max. permitted input voltage: 600V  

Warning

 

Risk of electric shock. 

 

When measuring resistance, be sure disconnect circuit. 

 

 

 