4.3.6. MIN/MAX and PEAK measurement
Fig. 9: MIN/MAX and PEAK measurement
1. Press key
MAX
W
to enter measurement of MAX and MIN values of the quantity to be
measured (see Fig. 9 - central part). The symbols
“
MAX
”
and
“
MIN
”
appear on the
display.
2. The values are automatically updated by the instrument whenever the currently
displayed values are exceeded (higher for MAX value, lower for MIN value).
3. Press key
HOLD/REL
to freeze the result. Message
“
HOLD
”
and
SAVE
S
appears on
the display
4. Press key
S
to save the value as a BMP image on the instrument
’
s internal memory or
press key
HOLD/REL
again to go back to function MAX/MIN
5. Press key
MAX
W
again or turn the selector switch to quit the function.
6. Press key
PEAK
X
to enter measurement of Peak values of the quantity to be
measured (see Fig. 9
–
right side). Symbols
“
Pmax
”
and
“
Pmin
”
appear on the display
and values are updated in the same way as for the MAX/MIN function.
7. Press key
HOLD/REL
to freeze the result. Message
“
HOLD
”
and
SAVE
S
appears on
the display
8. Press key
S
to save the value as a BMP image on the instrument
’
s internal memory or
press key
HOLD/REL
again to go back to the PEAK function.
9. Press key
PEAK
X
again or turn the selector switch to quit the function.
4.3.7. Detection of AC voltage without contact
CAUTION
x
Firstly use the NCV sensor on a known AC source in order to verify its
proper operation.
x
The thickness of the cable
’
s insulating sheath and the distance from the
source may influence the operation.
1. Switch on the instrument in any position of the selector switch
2. Take the instrument near an AC source and look for the red LED on the top to turn on
(see Fig. 1
–
part 3); this indicates that the instrument has detected the source's
presence