HT7052
EN - 18
6.4 VOLTAGE
MEASUREMENT
CAUTION
Maximum input for DC or AC voltage is 600V. Do not attempt to take any
voltage measurement that exceeds the limits. Exceeding the limits could
cause electrical shock and damage the instrument
1. Switch on the instrument by pressing the
ON/OFF
key
2. Select with arrow keys
or
the item “VOLTAGE” on main menu and confirm with
SELECT
key. The screen of Fig. 13 is shown by the meter
Fig. 13: Initial screen of voltage measure
Fig. 14: Screen of measured value
3. Connect the red part of the Test lead 1 or Test leads 2 (see § 4.2) to the
+OUTPUT
input and the black part of the Test leads 2 (see § 4.2) to the
–OUTPUT
input
4. Connect the tip of Test lead 1 or Test leads 2 (positive) and the black cable of Test
leads 2 (negative) to the object on test respect the polarities for DC voltage
measurement (see Fig. 15)
Fig. 15: Connection of meter for voltage measurement
5. Press
START/STOP
key to activate the measurement in continuous mode
6. Press again the
START/STOP
to stop the measurement. The result of test is shown at
display (see Fig. 14)
7. For saving the result see § 7