Power cable specifications (continued)
Cross-sectional area
4 mm² or 12 AWG
Terminals
Ferrule terminals, except the PE terminal, which uses the M6 ring type
External diameter range
8.5–14 mm
NOTE:
The main disconnect and branch protection should be provided by the installer.
Configuration 2: 200–240 V line-to-line one-phase configuration
Printer specifications
Number of power lines
2 lines + PE
Input voltage (line to line)
200–240 V~ (±10%)
Input frequency
50/60 Hz
Power consumption (typical)
2.6 kW
Maximum load current (per phase)
19 A
Branch circuit breaker specifications
2 poles, 25 A, type J
Power cable specifications
Configuration
2 lines + PE
Cross-sectional area
4 mm² or 12 AWG
Terminals
Ferrule terminals, except the PE terminal, which uses the M6 ring type
External diameter range
8.5–14 mm
Power cables
No power cable is provided with the processing station. The cables that you use must meet the minimum
specifications for the chosen configuration explained for each configuration.