Single Channel
Failures
A single channel failure, not caught by the self-test, can be caused by one of
two components: the channel plug-on module or the channel output relay.
To help isolate a plug-on module, disassemble the module and swap two
plug-on modules. DO NOT operate the module without a plug-on module
installed for every channel.
If a channel still fails after swapping the plug-on modules,suspect the
channel output relay.