Performance Verification Tests
These procedures test the electrical performance of the HP E1418A using
the specification in Appendix A as the performance standards. These tests
are suitable for incoming inspections, troubleshooting, and preventive
measures.
DC Voltage Output
Verification
This procedure tests the dc voltage output performance of the module.
1.
For each channel in the module, connect the 10 k
Ω
resistor and
DMM as shown in Figure 4-1. Channel 1 connections are shown in
the figure. Connect the other channels in a similar manner. Set the
DMM to measure dc volts.
2.
Set the channel to output each of the following voltages:
+16.00
– 4.00
+12.00
– 8.00
+ 8.00
–12.00
+ 4.00
–16.00
0.00
Use the
APPLy
n:VOLTage xx.xx
command, where
n
is the
channel number and
xx.xx
is the voltage to output.
3.
For each voltage, measure the output voltage with the DMM, enter it
in the table beginning on page 46, and compare the voltage output to
the limits shown in the table.
Figure 4-1. DC Voltage Output Connections (Channel 1)
34 Verification Tests
Chapter 4
Содержание E1418A
Страница 4: ...Notes 4 HP E1418A 8 16 Channel D A Converter Service Manual Contents ...
Страница 8: ...Notes 8 HP E1418A D A Converter Module Service Manual ...
Страница 58: ...Notes 58 Verification Tests Chapter 4 ...
Страница 66: ...Figure 5 5 Internal Resistor Adjustment Connections 66 Adjustments Chapter 5 ...
Страница 81: ...Figure 6 1 8 Channel Disassembly Chapter 6 Service 81 ...
Страница 82: ...Figure 6 2 16 Channel Disassembly 82 Service Chapter 6 ...
Страница 83: ...Figure 6 3 Plug on Channel Locations Figure 6 4 Installing Plug on Modules Chapter 6 Service 83 ...
Страница 84: ...Notes 84 Service Chapter 6 ...
Страница 90: ...Figure 7 1 HP E1418A Replaceable Parts 90 Replaceable Parts Chapter 7 ...
Страница 91: ...Figure 7 2 Common Terminal Module Replaceable Parts Chapter 7 Replaceable Parts 91 ...
Страница 94: ...Notes 94 Manual Changes Chapter 8 ...