Adjustment Procedure
Adjustment is performed on each channel, one at a time. Voltage output and
current output each use unique calibration constants.
Preparation
Before performing any adjustments, determine if the adjustments are to be
temporary or permanent (refer to the discussion on page 60).
If Permanent
1.
Turn off the mainframe.
2.
Remove the terminal module.
3.
Remove the module from the mainframe.
4.
Set the Cal Store Enable Jumper to the CAL position as shown below.
5.
Re-install the module in the mainframe.
6.
Make the CAL output terminal connections on the terminal module.
7.
Install the terminal module.
8.
Apply power and allow a 15 minute warm-up.
If Temporary
1.
Make the CAL output terminal connections on the terminal module.
2.
Install the terminal module.
3.
Apply power and allow a 15 minute warm-up.
Figure 5-2. Cal Store Enable Jumper
62 Adjustments
Chapter
5
Содержание E1418A
Страница 4: ...Notes 4 HP E1418A 8 16 Channel D A Converter Service Manual Contents ...
Страница 8: ...Notes 8 HP E1418A D A Converter Module Service Manual ...
Страница 58: ...Notes 58 Verification Tests Chapter 4 ...
Страница 66: ...Figure 5 5 Internal Resistor Adjustment Connections 66 Adjustments Chapter 5 ...
Страница 81: ...Figure 6 1 8 Channel Disassembly Chapter 6 Service 81 ...
Страница 82: ...Figure 6 2 16 Channel Disassembly 82 Service Chapter 6 ...
Страница 83: ...Figure 6 3 Plug on Channel Locations Figure 6 4 Installing Plug on Modules Chapter 6 Service 83 ...
Страница 84: ...Notes 84 Service Chapter 6 ...
Страница 90: ...Figure 7 1 HP E1418A Replaceable Parts 90 Replaceable Parts Chapter 7 ...
Страница 91: ...Figure 7 2 Common Terminal Module Replaceable Parts Chapter 7 Replaceable Parts 91 ...
Страница 94: ...Notes 94 Manual Changes Chapter 8 ...