Setting the Card ID Switch
The multiplexer module can be an HP E1351A, HP E1352A or HP E1353A
depending on which terminal module is installed. With a normal
connection, the terminal module will automatically configure the module
for the appropriate model. To identify a module without a terminal module,
set the Card ID switch as shown in Figure 2-6. The ID numbers are:
Model
Value
Configuration
HP E1351A
0
all pins open
HP E1352A
8
pin 3 closed
HP E1353A
2
pin 1 closed
For normal connections, leave the switch at the factory setting which is all
pins open.
Figure 2-6. Card ID Switch