Equivalent Front-Panel Key Commands
REF SET
The REF SET Data Function establishes a measurement reference point. The
HP-IB command :REFerence is used to select this data function
programmatically.
To Turn Measurement Reference Points ON and OFF.
Use the
:REFerence:STATe <boolean> commands to turn measurement reference points
ON and OFF. The <boolean> parameter can be a 1 or ON to turn measurement
reference points on, and a 0 or OFF to turn measurement reference points off.
Syntax
:REFerence:STATe <ON> or <1>
:REFerence:STATe <OFF> or <0>
Example
OUTPUT 714;"MEAS:RFR:POW:REF:STAT ON"
This turns the measurement reference point for the TX Power
measurement ON.
To Query the State of Measurement Reference Points.
Use the
:REFerence:STATe? commands to query the state of a measurement reference
point. The query returns a 1 if a measurement reference points is ON, and a 0 if a
measurement reference points is OFF.
Syntax
:REFerence:STATe?
Example
OUTPUT 714;"MEAS:RFR:POW:REF:STAT?"
ENTER 714;Meter_on_off ! returns a 1 (ON) or 0 (OFF)
This queries the state of the measurement reference point for the TX
Power measurement.
Содержание 8920B
Страница 10: ...10 ...
Страница 16: ...Contents 16 9 Error Messages Index 673 ...
Страница 17: ...17 1 Using HP IB ...
Страница 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Страница 49: ...49 2 Methods For Reading Measurement Results ...
Страница 61: ...61 3 HP IB Command Guidelines ...
Страница 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 85: ...85 4 HP IB Commands ...
Страница 90: ...90 Adjacent Channel Power ACP ...
Страница 116: ...116 Configure I O Configure ...
Страница 156: ...156 Call Processing ...
Страница 162: ...162 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Страница 242: ...242 Common Command Descriptions ...
Страница 251: ...251 5 Advanced Operations ...
Страница 377: ...377 7 IBASIC Controller ...
Страница 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 643: ...643 9 Error Messages ...
Страница 672: ...672 ...
Страница 682: ...Index 682 ...