81
Chapter 3, HP-IB Command Guidelines
Guidelines for Operation
Reading back the measurement, instrument function, or data function state
Use the query form of the command, STATe?, to determine the current state of a
measurement, instrument function or data function. If a measurement, instrument
function, or data function is queried, the returned value will be either a “1” (ON)
or a “0” (OFF).
For example, the following BASIC language statements illustrate the use of the
STATe? command to determine the current state of the TX Power measurement:
!Query the state of the TX Power measurement
OUTPUT 714;"MEAS:RFR:POW:STAT?"
ENTER 714;State_on_off
IF State_on_off = 1 THEN DISP "TX Power Measurement is ON"
IF State_on_off = 0 THEN DISP "TX Power Measurement is OFF
STATe Command Guidelines
•
Measurements that are displayed as numbers, or as analog meters using the METER
function, can be turned on and off.
•
The data functions REFerence, METer, HLIMit, and LLIMit can be turned on and off.
•
Any instrument function that generates a signal can be turned on and off. This includes
the RF Generator, Tracking Generator, AF Generator 1, AF Generator 2, and the
Signaling Encoder.
•
The Oscilloscope’s trace cannot be turned off.
•
The Spectrum Analyzer’s trace cannot be turned off.
Содержание 8920B
Страница 10: ...10 ...
Страница 16: ...Contents 16 9 Error Messages Index 673 ...
Страница 17: ...17 1 Using HP IB ...
Страница 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Страница 49: ...49 2 Methods For Reading Measurement Results ...
Страница 61: ...61 3 HP IB Command Guidelines ...
Страница 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 85: ...85 4 HP IB Commands ...
Страница 90: ...90 Adjacent Channel Power ACP ...
Страница 116: ...116 Configure I O Configure ...
Страница 156: ...156 Call Processing ...
Страница 162: ...162 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Страница 242: ...242 Common Command Descriptions ...
Страница 251: ...251 5 Advanced Operations ...
Страница 377: ...377 7 IBASIC Controller ...
Страница 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 643: ...643 9 Error Messages ...
Страница 672: ...672 ...
Страница 682: ...Index 682 ...