550
Chapter 8, Programming The Call Processing Subsystem
Programming The CALL BIT Screen
When the CALL BIT screen is displayed and the Call Processing Subsystem is in
the
Connect
state, the host firmware constantly monitors the mobile station’s
transmitted carrier power. If the power falls below 0.0005 Watts the error message
RF Power Loss indicates loss of Voice Channel
will be displayed
and the Test Set will terminate the call and return to the
Active
state.
In order to ensure that the host firmware makes the correct decisions regarding the pres-
ence of the mobile stations’s RF carrier, the Test Set’s RF power meter should be zeroed
before using the Call Processing Subsystem. Failure to zero the power meter can result in
erroneous RF power measurements. See
"Conditioning the Test Set for Call Process-
for information on zeroing the RF Power meter manually.
Refer to the User’s Guide for detailed information on the operation and manual
use of the
CALL BIT
screen. The field descriptions for each of the decoded
messages are given in the “CALL BIT Screen Message Field Descriptions”
section of “Call Processing Subsystem” chapter, in the User’s Guide.
The information presented in this section covers the
CALL BIT
screen
programming commands and how to use them.
Содержание 8920B
Страница 10: ...10 ...
Страница 16: ...Contents 16 9 Error Messages Index 673 ...
Страница 17: ...17 1 Using HP IB ...
Страница 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Страница 49: ...49 2 Methods For Reading Measurement Results ...
Страница 61: ...61 3 HP IB Command Guidelines ...
Страница 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 85: ...85 4 HP IB Commands ...
Страница 90: ...90 Adjacent Channel Power ACP ...
Страница 116: ...116 Configure I O Configure ...
Страница 156: ...156 Call Processing ...
Страница 162: ...162 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Страница 242: ...242 Common Command Descriptions ...
Страница 251: ...251 5 Advanced Operations ...
Страница 377: ...377 7 IBASIC Controller ...
Страница 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 643: ...643 9 Error Messages ...
Страница 672: ...672 ...
Страница 682: ...Index 682 ...