34
Remarks:
.Min Input Voltage
:
10V
·Allowable maximum overload protection voltage:600V RMS
·Allowable maximum overload protection current:1000A RMS
F.
Reactive Power (Var =V×A×SINФ)
Current/Voltage
Voltages Range
15V
100V
300V
600V
Current
Range
40A
0.60kVar
4.00kVar
12.00kVar
24.00kVar
100A
1.50kVar
10.00kVar
30.00kVar
60.00kVar
400A
6.00kVar
40.00kVar
120.0kVar
240.0kVar
1000A
15.00kVar
100.0kVar
300.0kVar
600.0kVar
Accuracy
15V/1000A Range :±(4%+25) Other
Range :±(4%+20)
Resolution
<
1000kVar:0.01kVar
≥ 100kVar:0.1kVar
Remarks
Min Input Current/Voltage
:
10A/10V
·Allowable maximum overload protection voltage:600V RMS
·Allowable maximum overload protection current:1000A RMS
G. Power Factor (PF = W/VA )
Range
Accuracy
Resolution
Measuring Condition
0.3~1
(capacitive or
inductive)
±0.022
0.001
The minimum measuring
current 10A
The minimum measuring
voltage 45V
0.3~1
(capacitive or
inductive)
For reference only
Measuring current less than
10A OR
Measuring voltage less than
45V
Remarks