Reflect Standards
Reflect Standards -Male and Female Short Circuits
Isolation Standards
Standards -Male and Female Broadband Loads
Precision Adapters
In this TRL 2-Port calibration, the reflect standards are
identical short circuits with one connected to Port 1 and
one to Port 2.
The male short circuit and the female short circuit are
manufactured to tolerances that assure equivalent
responses. If one is damaged or otherwise becomes
defective, calibration errors will result. Errors will also
result if the open circuits are used in the procedure instead
of the short circuits (unless the calibration kit standard
assignment definition is changed).
The fixed loads used for the isolation part of the calibration
are the same broadband loads as used for the O-2 loads
measurement. Exact equivalent responses are not required
for this step.
Three precision adapters are included in this kit. Two are
always recommended for the 3.5 mm TRL
measurement calibration. All three can be used in the
adapter removal calibration for measurement of
sertable devices. Complete performance verification
assumes use of the precision adapters in this kit as the test
ports.
MALE to FEMALE
MALE to MALE
FEMALE to FEMALE
Precision
test port adapters