Measure Standards
N
ow
perform the TRL 2-Port measurement calibration
procedure using the HP
3.5 mm Precision
Calibration Kit.
Note:
The standards are measured in the order: Thru,
Reflect, Line, Isolation in order to connect the load
standards just once.
Press:
CAL
CAL 1
TRL 2-PORT
l
Connect the TRL male adapter (Port 1) to the TRL female
adapter (Port 2).
l
Press THRU. The HP 8510 makes six measurements; they
are S,,, S,,, S,,, S,,, and two specially redefined user
parameters.
l
Connect the Short circuit to Port 1.
Press REFLECT SHORT.
is measured.
Reflect Short Standard
l
Connect the Short circuit to Port 2.
l
Press REFLECT SHORT. is measured.
l
Press LINES.
Reflect Short Standard
18