Inside the Sequence Operation State.
Figure
illustrates the
operation of the sequence operation state.
The downward entrance to the Sequence Operation State signals
that some instrument dependent action should begin at once. An
upward exit is not allowed until the instrument signals that its
action is complete. Note that complete can be defined differently for
different instruments. For example, consider an instrument that can
sweep a range of frequencies starting with and ending with
The
action-complete signal can be defined to coincide with the output of
either or
Figure
Inside the Sequence Operation State
Common Trigger
In the previous paragraphs, you learned about the basic building
Configurations
blocks allowed in a SCPI trigger system. Generally, an instrument
implements only a portion of the trigger features available. These
paragraphs discuss the simplest configurations:
and TRIG.
The
Configuration
The
configuration is the simplest possible trigger configuration.
It uses no event detection states, and requires only two subsystems
for programming,
and
.
All SCPI instruments
implement these two subsystems
Getting Started Programming
Содержание 8360
Страница 97: ......
Страница 138: ......
Страница 150: ......
Страница 259: ...HP 8360 User s Handbook FEEDFORWARD Figure M l ALC Block Diagram Operating and Programming Reference M 13 ...
Страница 318: ......
Страница 336: ......
Страница 352: ......
Страница 360: ......
Страница 397: ...2b Menu Maps Menu Maps 2b 1 ...
Страница 398: ......
Страница 399: ......
Страница 400: ......
Страница 401: ......
Страница 402: ......
Страница 403: ......
Страница 404: ......
Страница 405: ......
Страница 406: ......
Страница 435: ...Figure 3 3 Removing the Side StraPs and Feet NSTALLATION 3 ...
Страница 461: ......
Страница 463: ...5 4 Change B HP 8380 User s Handbook ...
Страница 467: ...5 8 Change A HP 8360 User s Handbook ...
Страница 481: ...optioll9l3RackFhgeKit 5 22 Change A HP 8360 User s Handbook ...