Servicing
the
HP
83447A
Adjustment
Procedures
Photodetector
assembly
adjustment
procedure
There
are
two
adjustments
on
the
photodetector
assembly
.
The
R26
+V
APD
adjust
sets
the
APD
bias
for
optimum
sensitivity
.
The
R1
High
Input
Limit
Set
sets
maximum
operating
power
for
jitter
performance
.
Since
the
equipment
setup
is
the
same
as
the
setup
required
for
testing
instrument
sensitivity
,
the
adjustment
procedures
are
included
after
the
performance
test.
Replacement
photodetector
assemblies
come
from
the
factory
pre-adjusted.
Slight
re-adjustments
may
be
required
as
the
instrument
ages
.
4-11
Содержание 83447A
Страница 1: ...User and Service Guide HP 83447A Lightwave Trigger Receiver ...
Страница 3: ...Printing History HP Part Number Edition Date 83447 90002 Edition 1 April 1996 iii ...
Страница 10: ......
Страница 15: ...1 General Information ...
Страница 25: ...Declaration of Conformity 1 11 ...
Страница 27: ...General Information Electrostatic Discharge Information Figure 1 3 Example of a static safe work station 1 13 ...
Страница 29: ...2 Installation and Preparation for Use ...
Страница 36: ...Installation and Preparation for Use Installing the HP 83447A Figure 2 3 AC power cables available 2 8 ...
Страница 51: ...3 Using the HP 83447A ...
Страница 55: ...4 Servicing the HP 83447A ...
Страница 93: ...Servicing the HP 83447A Replaceable parts Figure 4 8 HP 83447A assembly level replaceable parts 4 39 ...
Страница 95: ...Servicing the HP 83447A Replaceable parts Figure 4 9 HP 83447A replaceable hardware 4 41 ...
Страница 97: ...Index ...
Страница 102: ...W warranty 2 13 warranty information viii waveform test measuring eye diagram 3 3 Index 6 ...