Test Setup
HP
8116A
0
0
Oscilloscope
(HP 54503A)
0
H P 8112A
Figure 8-19. High-Level Control Verification Test
Procedure
1 . Connect the equipment as shown in the setup figure.
2. Set up the HP 81 12A as follows:
8-28 Testing Performance
Trigger Mode
Control Mode
Transition
PER
DEL
DTY
LOL
NORM
HILC
Fixed
200 µs
65 ns
50%
-2.00
v
3. Set up the HP 81 16A Pulse/Function Generator as follows:
Output
Sinewave
Frequency
50 Hz
Amplitude
16.0 V
Offset
0.00 V
4. Verify that the High-level output can be varied between +8 V and
-8 V while the -2 V level remains unchanged as shown below.
Control
Input
c���;:1:d
o v
-2
v
Figure 8-20. Correct High-level Control
5 . Set up the HP 8112A as follows:
Trigger Mode
NORM
Transition
Fixed
PER
10.0 ms
DEL
65 ns
WID
1 0 ns
LOL
-2.00 v
COMPL
On
Содержание 8112A
Страница 6: ......
Страница 24: ......
Страница 36: ......
Страница 44: ......
Страница 70: ......
Страница 92: ......
Страница 144: ...8 32 Testing Performance ...
Страница 176: ......
Страница 188: ...B END 1O 1 8 Instrument Overview and Troubleshooting Guide N E 3 5 N _ O E 3 1 r No E 5 r N E s 2 ...
Страница 190: ......
Страница 194: ...10 2 4 Servicing the Power Supply ...
Страница 196: ......
Страница 197: ......
Страница 216: ......
Страница 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Страница 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Страница 244: ......
Страница 251: ......
Страница 262: ...A B A2 ED RY CONTROL ...
Страница 270: ......
Страница 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Страница 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Страница 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Страница 334: ......
Страница 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Страница 379: ...A3 tlP3 HP40 4x P41 4x J4 i J2 I O a J I 1 I I a 4 I I I V3 tlP1 I Figure B 15 Mechanical Parts 2 Backdating B 45 ...
Страница 380: ......
Страница 382: ......
Страница 402: ......