Model 5305B
Operation
Figure 9E-3-3. Self-Check Measurements
●
Indicates overflow light is on.
9E-3-3
5300A Self Check 5300B Self Check
Resolution
Selector Display Annunciator Display Annunciator
10 kHz
0010.00
M, Hz
000010.00
±
1
M, Hz
1 kHz
010.000
M, Hz
00010.000
±
1
M, Hz
100 Hz
10.0000
M, Hz, C
0010.0000
±
1
M, Hz, C
10 Hz
0.00000
M, Hz, C
010.00000
±
1
M, Hz, C
1 Hz
●
000.000
K, Hz, C
10000.000
±
1
K, Hz, C
.1 Hz
●
00.0000
K, Hz, C
0000.0000
±
1
K, Hz, C
1.
Apply input power to 5300 ac receptacle.
2.
Turn counter on with 5300 SAMPLE RATE control.
Adjust SAMPLE RATE for minimum display time
(full ccw).
3.
Set RANGE switch to CHK position. Display is a
function of RESOLUTION switch.